The structural-geometric and functional parameters of the heart in patients after mitral prosthetics with new full-flow valves in the long-term period

Cover Page


Cite item

Full Text

Abstract

Aim. To compare the results of mitral valve replacement with the full-flow mechanical valve «MedInzh-ST» and the classical prosthesis «MedInzh-2», to learn the structural and functional parameters of the heart depending from the prosthesis in the long-tern period.

Material and methods. From 2015 to February 2020 years an the basis of the cardio-surgical department of the «Regional clinical Hospital» 116 patients, in order to correct defects, were underwent mitral prosthetics with «MedInzh» valves. Of these 55 patients received a new full-flow valve «MedInzh-ST». 61 patients – a classical «MedInzh-2». Before surgery and in the long-term period all patients were underwent transthoracic echocardiography with using expert-class devices. In the long-tern postoperative period 34 patients were underwent echocardiographic research after implantation full-flow valves and 40 patients after prosthetics of classical prosthesis.

Results. The choice of the prosthesis model did not affect the incidence of postoperative complications and the rate of hospital mortality. In all cause of death is not associated with impaired function of the prosthesis. All patients before the discharge from hospital were fixed the positive effect from surgical treatment. In the long-term postoperative period between patients after implantation of full-flow valves, the cases of prosthetic platelets were not registered. One patient was diagnosed with prosthetic thrombosis after implantation of a classical prosthesis, which required further reprosthetics. There were no signs of prosthetic endocarditis and paraprosthetic fistula in both groups. There were no significant differences in the structural and geometric parameters of the heart in all patients.

Conclusion. Analysis of the functional characteristics of the new full-flow valve «MedInzh-ST» in the long-term period allows us to conclude that the new valve model meets modern requirements for efficiency and safety.

Full Text

СПИСОК СОКРАЩЕНИЙ

ИК – искусственный клапан

КДО – конечный диастолический объем

КДР – конечный диастолический размер

КСО – конечный систолический объем

КСР – конечный систолический размер

ОНМК – острое нарушение мозгового кровообращения

ТМГД – трансмитральный градиент давления

Р ЛА – давление в легочной артерии

Введение

Пороки клапанов сердца составляют от 7 до 10% всех заболеваний сердца. Наиболее часто встречается поражение митрального клапана. Причинами митральных пороков являются дегенеративные изменения, ревматическая болезнь сердца, инфекционный эндокардит, дисплазия соединительной ткани сердца [1].

По данным хирургических регистров, в мире ежегодно имплантируется от 250 тыс. до 280 тыс. клапанных протезов сердца. Несмотря на очевидный прогресс в хирургии сердца, совершенствование реконструктивно-пластических или эндоваскулярных транскатетерных операций, число имплантаций искусственных клапанов (ИК) не снижается. Протезирование клапанов сердца остается эффективным и часто единственно возможным способом коррекции пороков, позволяющим устранить патологические изменения, улучшить внутрисердечную гемодинамику и качество жизни пациентов [2, 3].

В настоящее время не существует идеальной модели ИК сердца. Сегодня приоритетным направлением в создании ИК сердца является разработка протезов с максимально возможным эффективным отверстием для обеспечения гемодинамики, приближенной к нативному клапану.

Новой усовершенствованной моделью ИК является отечественный полнопроточный двустворчатый клапан «МедИнж-СТ». Главным преимуществом данного протеза является его конструкция: створки фиксированы на шарнирных креплениях, находящихся на противоположных сторонах кольца, что способствует устранению застойных зон вокруг креплений и снижает вероятность развития тромботических осложнений. Запирающий элемент, выполненный в виде двух цилиндрических сегментов, обеспечивает централизацию потока крови, минимальную травматизацию форменных элементов, увеличение эффективной площади отверстия клапана и уменьшение транспротезного градиента давления. Прочность клапана «МедИнж-СТ» обусловлена исходной высокой прочностью монолитного изотропного пиролитического углерода [2, 4].

Цель исследования – сравнить результаты протезирования митрального клапана полнопроточными механическими протезами «МедИнж-СТ» и классическими протезами «МедИнж-2», изучить структурно-геометрические и функциональные параметры сердца в зависимости от модели протеза в отдаленном периоде.

Материал и методы

С 2015 по февраль 2020 г. на базе кардиохирургического отделения БУЗОО «Областная клиническая больница» 116 пациентам с целью коррекции пороков было выполнено митральное протезирование клапанами «МедИнж». До включения в исследование от всех участников было получено письменное информированное согласие. В исследование включены все пациенты, перенесшие коррекцию митрального порока механическими протезами «МедИнж». Пациентов с ишемической болезнью сердца и перенесенным инфарктом миокарда в исследование не включали. Операция выполнена по стандартной методике в условиях искусственного кровообращения, нормотермии.

Условий и показаний для имплантации той или иной модели протеза для определенных групп пациентов не было, выбор модели клапана зависел от материально-технического обеспечения. Новый полнопроточный протез «МедИнж-СТ» имплантирован 55 пациентам (группа 1), классический «МедИнж-2» – 61 пациенту (группа 2). В группе 1 преобладали пациенты мужского пола, в группе 2 – женского. Средний возраст пациентов составил 57 (50; 62) и 55 (50; 62) лет соответственно.

В отдаленном послеоперационном периоде эхокардиографическое исследование выполнено 34 пациентам после имплантации полнопроточных клапанов и 40 пациентам – после протезирования классическими протезами. Средняя продолжительность наблюдения в группе 1 составила 2,8 [1; 3] года, в группе 2 – 3,1 [2; 3, 5] года. Данный период наблюдения является достаточным для оценки процессов ремоделирования сердца после операции и выявления поздних клапанзависимых послеоперационных осложнений.

Всем пациентам перед операцией и в отдаленном периоде выполняли трансторакальную эхокардиографию с использованием аппаратов экспертного класса Vivid E9, GE Vivid Q General Electric (США). Определяли следующие параметры: конечный диастолический размер (КДР), конечный систолический размер (КСР), конечный диастолический объем (КДО), конечный систолический объем (КСО), фракцию выброса левого желудочка, размер левого предсердия, пиковый и средний трансмитральные градиенты давления (ТМГД), степень митральной регургитации, размер правого желудочка, систолическое давление в легочной артерии (Р ЛА). Существенных различий дооперационных эхокардиографических показателей среди лиц, перенесших протезирование клапанами разных моделей, не было.

Статистические методы. Для обработки информации использовались программы Statistica 12, Microsoft Excel. Вид распределения вариационных рядов оценивался при помощи критерия Шапиро–Уилка. Вид распределения отличался от нормального, в связи с чем анализ проводили с помощью непараметрических методов. Для описания количественных данных рассчитывались медиана (Me) и процентили (Р25, Р75), для качественных данных – доли. Сравнение количественных данных двух независимых выборок осуществлялось при помощи U-критерия Манна–Уитни, двух зависимых выборок – критерия Уилкоксона. Для анализа различия частот в двух независимых группах использовался критерий χ2 с поправкой Йейтса. Статистическая значимость присваивалась при значении p<0,05.

Результаты

Причинами пороков сердца были ревматическая болезнь сердца, инфекционный эндокардит, дисплазия соединительной ткани сердца, в том числе миксоматозная дегенерация. Для определения влияния модели протеза на послеоперационные показатели пациенты каждой группы были разделены на подгруппы в зависимости от вида порока: преобладание стеноза или недостаточности. В группе 1 митральный стеноз преобладал у 12 пациентов (группа 1.1), в группе 2 – у 16 пациентов (группа 2.1). Митральная недостаточность преобладала у 43 пациентов группы 1 (группа 2.1), у 45 пациентов группы 2 (группа 2.2). Все пациенты были сопоставимы по клиническим и эхокардиографическим характеристикам (табл. 1, 2).

 

Таблица 1. Дооперационные эхокардиографические параметры у пациентов обеих групп

Table 1. Preoperative echocardiographic parameters in patients of both groups

Параметр

Группа 1.1 (n=12)

Группа 2.1 (n=16)

р1.1–2.1

Группа 1.2 (n=43)

Группа 2.2 (n=45)

р1.2–2.2

КДО, мл

106 [98; 124]

95 [91; 114]

0,186

153 [147; 161]

150 [144; 159]

0,740

КСО, мл

38 [31; 42]

41 [38; 54]

0,821

57 [50; 62]

60 [55; 67]

0,864

КДР, см

4,9 [4, 3; 5, 1]

4,7 [4, 2; 4, 9]

0,746

5,7 [5, 3; 5, 9]

5,6 [5, 2; 5, 9]

0,896

КСР, см

3,2 [3, 0; 3, 5]

3,3 [3, 2; 3, 5]

0,687

3,6 [3, 2; 3, 8]

3,7 [3, 2; 4, 0]

0,862

ФВ, %

63 [55; 68]

61 [56; 70]

0,673

64 [57; 69]

62 [58; 72]

0,716

ЛП, см

5,1 [4, 8; 5, 3]

4,8 [4, 5; 5, 1]

0,098

4,6 [4, 4; 5, 1]

4,8 [4, 5; 5, 2]

0,520

пикТМГД, мм рт. ст.

25 [18; 28]

27 [22; 31]

0,616

7 [5; 8]

8 [4; 10]

0,614

срТМГД, мм рт. ст.

14 [12; 18]

14 [11; 18]

0,842

4,0 [3, 0; 5, 5]

6,0 [4, 0; 8, 0]

0,738

Степень регургитации

2,5 [2, 0; 3, 0]

2,0 [2, 0; 3, 0]

0,549

3 [3; 4]

3 [3; 4]

0,898

ПЖ, см

3,3 [3, 1; 3, 6]

2,7 [2, 6; 3, 0]

0,087

2,6 [2, 3; 3, 1]

2,7 [2, 3; 3, 0]

0,765

Р ЛА, мм рт. ст.

67 [62; 71]

57 [54; 63]

0,404

45 [38; 51]

55 [46; 60]

0,343

Примечание. Здесь и в табл. 3–6: ФВ ЛЖ – фракция выброса левого желудочка, ЛП – размер левого предсердия, пикТМГД – пиковый трансмитральный градиент давления, срТМГД – средний трансмитральный градиент давления, ПЖ – правый желудочек.

 

Таблица 2. Клиническая характеристика пациентов

Table 2. Clinical characteristics of patients

Параметр

Группа 1 (n=55)

Группа 2 (n=61)

р

Мужской пол, абс. (%)

32 (58)

22 (36)

0,028

Возраст, лет, Ме (Р25; Р75)

57 (50; 62)

55 (50; 62)

0,832

ИМТ, кг/м2, Ме (Р25; Р75)

27,4 (23,1; 30,9)

27,6 (24,5; 32,2)

0,817

Избыточная масса тела, абс. (%)

18 (33)

15 (25)

0,445

Ожирение, абс. (%)

16 (29)

17 (28)

0,952

Гипертоническая болезнь, абс. (%)

28 (51)

30 (49)

0,853

Сахарный диабет, абс. (%)

6 (11)

4 (7)

0,616

ХОБЛ, абс. (%)

19 (35)

11 (18)

0,070

Фибрилляция предсердий, абс. (%)

20 (36)

23 (38)

0,966

Давность ухудшения, месяцев, Ме (Р25; Р75)

8 (6; 14)

6 (3; 12)

0,656

НК по NYHA, Ме (Р25; Р75)

3 (2,5; 3)

3 (2,7; 3)

0,846

Ревматическая болезнь сердца, абс. (%)

15 (27)

21 (34)

0,529

Миксоматозная дегенерация без отрыва хорд, абс. (%)

12 (22)

8 (13)

0,321

ИЭ без отрыва хорд, абс. (%)

4 (7)

8 (13)

0,468

ДСТ без отрыва хорд, абс. (%)

4 (7)

6 (10)

0,873

Отрыв хорд, абс. (%)

19 (35)

18 (30)

0,703

Тромбоз протеза, абс. (%)

1 (2)

0

0,431

Преобладание стеноза, абс. (%)

12 (22)

16 (26)

0,737

Преобладание недостаточности, абс. (%)

43 (78)

45 (74)

0,737

Повторное протезирование, абс. (%)

2 (4)

4 (7)

0,773

Примечание. ДСТ – дисплазия соединительной ткани, ИМТ – индекс массы тела, ИЭ – инфекционный эндокардит, НК по NYHA – недостаточность кровообращения по NYHA, ХОБЛ – хроническая обструктивная болезнь легких.

 

Интраоперационная летальность в обеих группах не зарегистрирована.

За время стационарного лечения в группе 1 зарегистрирован один случай тромбоза протеза у пациентки с доказанной тромбофилией, в группе 2 случаев тромбоза протеза не было. У пациентов после протезирования полнопроточными клапанами зарегистрирован 1 случай транзиторной ишемической атаки, в группе классических протезов – 2 случая острого нарушения мозгового кровообращения (ОНМК) по ишемическому типу, один случай летального ОНМК по геморрагическому типу с образованием внутримозговой гематомы.

В 30-дневный период после операции в группе полнопроточных протезов зафиксировано 2 (3,6%) случая летального исхода, в группе классических протезов – 5 (8,2%) случаев (р=0,523). Во всех случаях смерть не была связана с нарушением функции протеза.

Результаты эхокардиографического исследования перед выпиской подтверждают эффективность оперативного лечения в обеих группах (табл. 3, 4). Нарушений в работе протеза не зарегистрировано.

 

Таблица 3. Послеоперационные эхокардиографические параметры у пациентов группы 1

Table 3. Postoperative echocardiographic parameters in group 1 patients

Параметр

Стеноз

Недостаточность

р1–2; 3–4

до операции (1) (n=12)

после операции (2) (n=12)

до операции (3) (n=43)

после операции (4) (n=41)

КДО, мл

106 [98; 124]

124 [112; 135]

153 [147; 161]

117 [110;125]

0,364; 0,036

КСО, мл

38 [31; 42]

55 [49; 64]

57 [50; 62]

45 [40; 49]

0,085; 0,028

КДР, см

4,9 [4, 3; 5, 1]

5,1 [4, 8; 5, 5]

5,7 [5, 3; 5, 9]

5,0 [4, 7; 5, 2]

0,246; 0,042

КСР, см

3,2 [3, 0; 3, 5]

3,6 [3, 2; 3, 9]

3,6 [3, 2; 3, 8]

3,3 [3, 2; 3, 5]

0,482; 0,034

ФВ ЛЖ, %

63 [55; 68]

58 [55; 60]

64 [57; 69]

60 [57; 64]

0,670; 0,267

ЛП, см

5,1 [4, 8; 5, 3]

4,7 [4, 5; 4, 9]

4,6 [4, 4; 5, 1]

4,3 [4, 0; 4, 5]

0,095; 0,040

пикТМГД, мм рт. ст.

25 [18; 28]

13 [11; 15]

7 [5; 8]

9 [7; 12]

0,028; 0,568

срТМГД, мм рт. ст.

14 [12; 18]

5 [4; 8]

4 [3; 5, 5]

5 [4; 7]

0,009; 0,612

Степень регургитации

2,5 [2, 0; 3, 0]

1,3 [1, 0; 1, 5]

3 [3; 4]

1,4 [1, 0; 1, 7]

0,043; 0,033

ПЖ, см

3,3 [3, 1; 3, 6]

3,1 [2, 9; 3, 3]

2,6 [2, 3; 3, 1]

2,6 [2, 4; 3, 0]

0,045; 0,864

Р ЛА, мм рт. ст.

67 [62; 71]

48 [45; 54]

45 [38; 51]

31 [28; 37]

0,027; 0,038

 

Таблица 4. Послеоперационные эхокардиографические параметры у пациентов группы 2

Table 4. Postoperative echocardiographic parameters in group 2 patients

Параметр

Стеноз

Недостаточность

р1–2; 3–4

до операции (1) (n=16)

после операции (2) (n=14)

до операции (3) (n=45)

после операции (4) (n=42)

КДО, мл

95 [91; 114]

107 [98; 114]

150 [144; 159]

131 [122; 140]

0,504; 0,028

КСО, мл

41 [38; 54]

44 [39; 52]

60 [55; 67]

54 [49; 57]

0,412; 0,030

КДР, см

4,7 [4, 2; 4, 9]

4,8 [4, 5; 5, 0]

5,6 [5, 2; 5, 9]

5,3 [4, 9; 5, 5]

0,798; 0,041

КСР, см

3,3 [3, 2; 3, 5]

3,3 [3, 0; 3, 5]

3,7 [3, 2; 4, 0]

3,6 [3, 4; 3, 9]

0,784; 0,562

ФВ, %

61 [56; 70]

58 [54; 60]

62 [58; 72]

59 [53; 62]

0,435; 0,238

ЛП, см

4,8 [4, 5; 5, 1]

4,7 [4, 4; 4, 8]

4,8 [4, 5; 5, 0]

4,5 [4, 2; 4, 8]

0,187; 0,034

пикТМГД, мм рт. ст.

27 [22; 31]

13 [10; 14]

8 [4; 10]

11 [9; 14]

0,008; 0,095

срТМГД, мм рт. ст.

14 [11; 18]

4 [3, 5; 6, 0]

6 [4; 8]

5 [4; 10]

0,012; 0,180

Степень регургитации

2 [1, 5; 2, 0]

1,5 [1, 0; 2, 0]

3 [3; 4]

1,8 [1, 5; 2, 0]

0,403; 0,019

ПЖ, см

2,7 [2, 6; 3, 0]

2,6 [2, 5; 2, 7]

2,7 [2, 3; 3, 0]

2,5 [2, 2; 2, 8]

0,346; 0,204

Р ЛА, мм рт. ст.

57 [54; 63]

38 [32; 43]

55 [46; 60]

34 [30; 42]

0,036; 0,012

 

Эхокардиографические параметры сердца перед выпиской были сопоставимы у пациентов вне зависимости от имплантируемой модели протеза (табл. 5).

 

Таблица 5. Послеоперационные эхокардиографические параметры у пациентов обеих групп

Table 5. Postoperative echocardiographic parameters in patients of both groups

Параметр

Группа 1.1 (n=12)

Группа 2.1 (n=14)

р1.1–2.1

Группа 1.2 (n=41)

Группа 2.2 (n=42)

р1.2–2.2

КДО, мл

106 [98; 124]

95 [91; 114]

0,186

153 [147; 161]

150 [144; 159]

0,740

КСО, мл

38 [31; 42]

41 [38; 54]

0,821

57 [50; 62]

60 [55; 67]

0,864

КДР, см

4,9 [4, 3; 5, 1]

4,7 [4, 2; 4, 9]

0,746

5,7 [5, 3; 5, 9]

5,6 [5, 2; 5, 9]

0,896

КСР, см

3,2 [3, 0; 3, 5]

3,3 [3, 2; 3, 5]

0,687

3,6 [3, 2; 3, 8]

3,7 [3, 2; 4, 0]

0,862

ФВ, %

63 [55; 68]

61 [56; 70]

0,673

64 [57; 69]

62 [58; 72]

0,716

ЛП, см

5,1 [4, 8; 5, 3]

4,8 [4, 5; 5, 1]

0,098

4,6 [4, 4; 5, 1]

4,8 [4, 5; 5, 2]

0,520

пикТМГД, мм рт. ст.

25 [18; 28]

27 [22; 31]

0,616

7 [5; 8]

8 [4; 10]

0,614

срТМГД, мм рт. ст.

14 [12; 18]

14 [11; 18]

0,842

4,0 [3, 0; 5, 5]

6,0 [4, 0; 8, 0]

0,738

Степень регургитации

2,5 [2, 0; 3, 0]

2,0 [2, 0; 3, 0]

0,549

3 [3; 4]

3 [3; 4]

0,898

ПЖ, см

3,3 [3, 1; 3, 6]

2,7 [2, 6; 3, 0]

0,087

2,6 [2, 3; 3, 1]

2,7 [2, 3; 3, 0]

0,765

Р ЛА, мм рт. ст.

67 [62; 71]

57 [54; 63]

0,404

45 [38; 51]

55 [46; 60]

0,343

 

В отдаленном периоде среди пациентов после имплантации полнопроточными клапанами случаев тромбоза протеза не зарегистрировано. У 1 пациентки после имплантации классического протеза выявлен тромбоз протеза, потребовавший в дальнейшем репротезирования. При этом все пациенты имели уровень МНО в терапевтическом диапазоне. В обеих группах признаков протезного эндокардита и парапротезной фистулы не выявлено. Значимых различий структурно-геометрических показателей сердца у всех исследуемых выявлено не было (табл. 6). Средний транспротезный градиент давления в обеих группах составил 4,4 [3, 4; 5, 4] и 4,7 [3, 4; 5, 5] мм рт. ст. соответственно.

 

Таблица 6. Эхокардиографические параметры у пациентов после протезирования митрального клапана в зависимости от модели протеза в отдаленном периоде

Table 6. Echocardiographic parameters in patients after mitral valve replacement depending on the prosthesis model in the long-term period

Параметр

МедИнж-СТ

МедИнж-2

р

КДО, мл

121 [101; 127]

115 [98; 143]

0,728

КСО, мл

42 [38; 53]

43 [35; 52]

0,960

КДР, см

5,1 [4, 7; 5, 2]

5,0 [4, 6; 5, 5]

0,857

КСР, см

3,3 [3, 1; 3, 5]

3,3 [3, 0; 3, 5]

0,983

ФВ, %

64 [60; 69]

63 [59; 66]

0,846

ЛП, см

4,6 [4, 2; 5, 3]

4,6 [4, 2; 5, 3]

0,986

пикТМГД, мм рт. ст.

10,9 [8, 2; 12, 7]

11,1 [8, 9; 13, 9]

0,581

срТМГД, мм рт. ст.

4,4 [3, 4; 5, 4]

4,7 [3, 4; 5, 5]

0,425

Степень регургитации

1,5 [1, 0; 2, 0]

2,0 [1, 5; 2, 0]

0,389

ПЖ, см

2,6 [2, 3; 2, 9]

2,6 [2, 3; 2, 9]

0,873

Р ЛА, мм рт. ст.

34 [26; 43]

34 [28; 42]

0,716

 

Обсуждение

В современной литературе данные о частоте послеоперационных осложнений у пациентов, перенесших протезирование митрального клапана отечественным двустворчатым протезом «МедИнж-2», сопоставимы с результатами зарубежных исследований по изучению иностранных моделей клапанов [1, 5, 6]. В настоящее время отсутствуют данные о динамике структурно- геометрических и функциональных параметров сердца после имплантации протеза «МедИнж-СТ» в митральную позицию. В настоящем исследовании мы оценили ранние и отдаленные эхокардиографические показатели сердца после хирургического лечения митральных пороков протезами «МедИнж».

Выбор модели протеза не влиял на показатель госпитальной летальности: в группе полнопроточных протезов – 3,6%, в группе классических протезов – 8,2%. Согласно изученной литературе, показатели летальности в исследуемых группах существенно не отличались от показателей после митрального протезирования другими моделями клапанов [1, 4, 5, 7].

Анализ структурно-функциональных параметров сердца перед выпиской из стационара подтвердил эффективность оперативного лечения в обеих группах. У всех пациентов с преобладанием митрального стеноза наблюдалось значимое снижение величины пикового и среднего ТМГД, систолического давления в легочной артерии. У пациентов с преобладанием митральной недостаточности после операции зарегистрировано значимое уменьшение КДО, КСР, КДР левого желудочка, размеров левого предсердия, Р ЛА и степени регургитации. Динамика данных показателей не зависела от выбранной модели протеза и была сопоставима с показателями после имплантации других моделей клапанов [1, 4, 7]. У всех исследуемых к моменту выписки нарушений функции протеза зарегистрировано не было.

Среди пациентов, перенесших коррекцию порока полнопроточными клапанами, к моменту выписки при эхокардиографическом исследовании выявлено достоверное снижение размеров правого желудочка и степени регургитации в случаях преобладания митрального стеноза, а также значимое уменьшение КСР левого желудочка в случаях преобладания митральной недостаточности. Выявленная динамика параметров сердца свидетельствует о благоприятном влиянии полнопроточных протезов на обратное ремоделирование сердца, чего не было выявлено среди пациентов после протезирования классическими клапанами «МедИнж-2».

Данные эхокардиографических исследований пациентов обеих групп в отдаленном послеоперационном периоде подтвердили положительную динамику структурно-геометрических параметров сердца. Величина пикового и среднего ТМГД была в пределах допустимых значений: 4,4 [3, 4; 5, 4] (группа 1) и 4,7 [3, 4; 5, 5] мм рт. ст. (группа 2). Нарушений функции протезов не выявлено. Значимых различий структурно-функциональных параметров сердца и функциональных характеристик клапанов в зависимости от модели протеза не получено. Отдаленные послеоперационные результаты были сопоставимы с показателями после оперативного лечения другими моделями клапанов [4–8], что подтверждает эффективность данных протезов при коррекции митральных пороков.

Заключение

Анализ структурно-функциональных параметров сердца в послеоперационном периоде показал, что полнопроточный клапан «МедИнж-СТ» в большей степени влияет на обратное ремоделирование правого желудочка и отвечает современным требованиям эффективности и безопасности. Функциональные характеристики клапанов обеих моделей (средний и пиковый ТМГД) оказались сопоставимыми.

В группе пациентов, перенесших протезирование полнопроточным клапаном, не было случаев тромбоза протеза ни в раннем, ни в отдаленном периоде. Таким образом, клапан «МедИнж-СТ» подтвердил свою эффективность при коррекции митральных пороков.

Ограничения исследования. Основными ограничениями исследования являются отсутствие рандомизации при разделении на группы и объем выборки.

Раскрытие интересов. Авторы декларируют отсутствие явных и потенциальных конфликтов интересов, связанных с публикацией настоящей статьи.

Disclosure of interest. The authors declare that they have no competing interests.

Вклад авторов. Авторы декларируют соответствие своего авторства международным критериям ICMJE. Все авторы в равной степени участвовали в подготовке публикации: разработка концепции статьи, получение и анализ фактических данных, написание и редактирование текста статьи, проверка и утверждение текста статьи.

Authors’ contribution. The authors declare the compliance of their authorship according to the international ICMJE criteria. All authors made a substantial contribution to the conception of the work, acquisition, analysis, interpretation of data for the work, drafting and revising the work, final approval of the version to be published and agree to be accountable for all aspects of the work.

Источник финансирования. Авторы декларируют отсутствие внешнего финансирования для проведения исследования и публикации статьи.

Funding source. The author declares that there is no external funding for the exploration and analysis work.

×

About the authors

Veronika S. Perekopskaya

Omsk State Medical University; Regional Clinical Hospital

Email: perekopskaya.vs@yandex.ru
ORCID iD: 0000-0002-6195-9747

cardiologist, Regional Clinical Hospital, Graduate Student

Russian Federation, Omsk; Omsk

Nataliya A. Morova

Omsk State Medical University

Email: nataliya-morova@yandex.ru
ORCID iD: 0000-0003-0003-692X

D. Sci. (Med.), Prof.

Russian Federation, Omsk

Valeriy N. Tsekhanovich

Omsk State Medical University; Regional Clinical Hospital

Author for correspondence.
Email: cvn50omsk@gmail.com
ORCID iD: 0000-0001-8300-1348

D. Sci. (Med.), Prof.

Russian Federation, Omsk; Omsk

References

  1. Маркова М.М., Тарасов Д.Г., Чернов И.И., и др. Фатальные и клинически значимые осложнения отдаленного периода после протезирования митрального клапана механическими протезами. Клиническая и экспериментальная хирургия. 2018;6(3):70-6 [Markova MM, Tarasov DG, Chernov II, et al. Fatal and moderate-to-severe long-term complications after mitral valve replacement. Clin Experiment Surg. 2018;6(3):70-6 (in Russian)]. doi: 10.24411/2308-1198-2018-13007
  2. Иванов В.А., Евсеев Е.П., Айдамиров Я.А., и др. Эволюция протезирования митрального клапана. Хирургия. Журнал им. Н.И. Пирогова. 2018;7:23-8 [Ivanov VA, Evseev EP, Aidamirov YaA, et al. Evolution of mitral valve replacement. Khirurgiya. Zhurnal im. N.I. Pirogova. 2018;7:23-8 (in Russian)]. doi: 10.17116/hirurgia2018723
  3. Кузьмина О.К., Рутковская Н.В. Ремоделирование миокарда при поражениях клапанов сердца. Сибирское медицинское обозрение. 2017;104(2):5-14 [Kuz'mina OK, Rutkovskaya NV. Myocardium remodeling in the heart valves disorders. Sibirskoe medicinskoe obozrenie. 2017;104(2):5-14 (in Russian)]. doi: 10.20333/2500136-2017-2-5-14
  4. Абдульянов И.В., Вагизов И.И., Каипов А.Э. Клинические результаты протезирования клапанов сердца двустворчатым полнопроточным механическим протезом «МедИнж-СТ». Ангиология и сосудистая хирургия. 2020;26(4):141-8 [Abdul'yanov IV, Vagizov II, Kaipov AE. Clinical results of prosthetics of heart valves with a bivalve full-flow mechanical prosthesis «MedInzh-ST». Angiology and vascular surgery. 2020;26(4):141-8 (in Russian)]. doi: 10.33529/ANGIO2020419
  5. Coutinho GF, Correia PM, Branco C, et al. Long-term results of mitral valve surgery for degenerative anterior leaflet or bileaflet prolapse: analysis of negative factors for repair, early and late failures, and survival. Eur J Cardiothorac Surg. 2016;50(1):66-74. doi: 10.15761/CDM.1000153
  6. Пукас Е.В., Лазоришинец В.В. Отдаленные результаты протезирования митрального клапана. Украинский журнал сердечно- сосудистой хирургии. 2018;10:44-7 [Pukas EV, Lazorishinec VV. Long-term results of mitral valve replacement. Ukrainian Journal of Cardiovascular Surgery. 2018;10:44-7 (in Russian)]. doi: 10.30702/ujcvs/18.32/10(044-047)
  7. Мартьянова Ю.Б., Кондратьев Д.А., Маркова М.М., и др. Непосредственные результаты хирургического лечения дисплазий митрального клапана. Клиническая и экспериментальная хирургия. 2019;7(1):28-33. [Mart'yanova YuB, Kondrat'ev DA, Markova MM, et al. Fatal and moderate-to-severe long-term complications after mitral valve replacement. Clinical and Experimental Surgery. 2019;7(1):28-33 (in Russian)]. doi: 10.24411/2308-1198-2019-11004
  8. Ивлева О.В., Авдеева М.В. Влияние протезирования митрального клапана на функциональную перестройку сердца у больных с синусовым ритмом и фибрилляцией предсердий. Креативная кардиология. 2018;12(1):40-9 [Ivleva OV, Avdeeva MV. Functional heart restoration after mitral valve replacement in patients with sinus rhythm and atrial fibrillationj. Creative Cardiology. 2018;12(1):40-9 (in Russian)]. doi: 10.24022/1997-3187-2018-12-1-40-49

Supplementary files

Supplementary Files
Action
1. JATS XML

Copyright (c) 2022 Eco-Vector

Creative Commons License
This work is licensed under a Creative Commons Attribution-NonCommercial-ShareAlike 4.0 International License.

СМИ зарегистрировано Федеральной службой по надзору в сфере связи, информационных технологий и массовых коммуникаций (Роскомнадзор).
Регистрационный номер и дата принятия решения о регистрации СМИ: ПИ № ФС 77 - 64546 от 22.01.2016. 


This website uses cookies

You consent to our cookies if you continue to use our website.

About Cookies